* Percentage of specimens that were repeatedly reactive.
Two of three specimens were also antibody positive. The third specimen was from a donor who subsequently showed no other evidence of HIV infection. The specificity of Manufacturer As antigen assay is approximately 99.95% (301,555 per 301,697) based on a) an assumed zero prevalence of HIV-1 p24 antigen in random blood donors who have no other evidence of HIV-1 infection and b) additional testing using the neutralization test and Western blot analysis for HIV-1 antibodies.
§ The repeatedly reactive specimen was confirmed as being HIV-1 p24-antigen positive; the donor subsequently seroconverted to antibodies to HIV-1. When the HIV-1 p24-antigenpositive donor was removed from the calculation, the specificity of Manufacturer Bs test was 99.93% (10,262 per 10,269; 95% confidence interval=99.86%99.97%).